Abstract
This study aimed to assess grid alignment for radiology departments in Khartoum state.
A total of 10 x-ray units in radiology departments in Khartoum state from different manufacture were tested by grid alignment test tool and the optical density for radiographic film was read by transition densitometer .
In this study the final result showed that the percentage of grid misalignment for x-ray machines was (40%), less than those with grid alignment x-ray machines (60%).
Which means almost of the grid were found misalignment. The study recommended that all x-ray equipment should be regularly checked using quality assurance tests.
Fadlalmola, R (2021). Assessment of Grid Alignment of X-Ray Units in Khartoum State.. Afribary. Retrieved from https://track.afribary.com/works/assessment-of-grid-alignment-of-x-ray-units-in-khartoum-state
Fadlalmola, Rawda "Assessment of Grid Alignment of X-Ray Units in Khartoum State." Afribary. Afribary, 21 May. 2021, https://track.afribary.com/works/assessment-of-grid-alignment-of-x-ray-units-in-khartoum-state. Accessed 27 Nov. 2024.
Fadlalmola, Rawda . "Assessment of Grid Alignment of X-Ray Units in Khartoum State.". Afribary, Afribary, 21 May. 2021. Web. 27 Nov. 2024. < https://track.afribary.com/works/assessment-of-grid-alignment-of-x-ray-units-in-khartoum-state >.
Fadlalmola, Rawda . "Assessment of Grid Alignment of X-Ray Units in Khartoum State." Afribary (2021). Accessed November 27, 2024. https://track.afribary.com/works/assessment-of-grid-alignment-of-x-ray-units-in-khartoum-state