Abstract
The purpose of this study to assessment radiation dose of CT scanner on
brain by calculate effective dose , then to confirm is there any difference
between effective dose of male brain and female brain according to
differences at anatomical structure of male brain and female brain , and to
determine typical parameter that help to reduce overdose .
This study performed in Modern Medical Center , the study showed
amount of average effective dose (1.26756002 mSv ± 0.2 mSv ) and it's
not equal or above than standard effective dose of brain (2 mSv) that
recommended by ICRP . Also there is no any difference between
effective dose of male brain and female brain according to differences at
anatomical structure of brain .The typical parameter of (Kvp and mAs ) is
(120 Kvp) and (60 mAs -100 mAs) for adult and pediatric