Abstract
This study evaluate 27 X-ray units worked in the Primary Health Care
Centers in Khartoum State, these units divided in three groups according to
the manufacturing company, the first group acts as 66.7% of units under
study it has mechanical problems which make the move of unit too difficult
also there is a problem in kilovolt and time accuracy .The second group acts
as 22.2% of all units which has programming and software problem. The
third one acts as 11.1% of all units which has no problems compared with
the previous groups. For the purchasing operation there is no obvious
system organized it, also there is no quality control program and team or person who is responsible or concerned with quality control tests
Rahman, A (2021). Evaluation Of Conventional X-Ray Machines In The Primary Health Care Centers In Khartoum State. Afribary. Retrieved from https://track.afribary.com/works/evaluation-of-conventional-x-ray-machines-in-the-primary-health-care-centers-in-khartoum-state
Rahman, Amal "Evaluation Of Conventional X-Ray Machines In The Primary Health Care Centers In Khartoum State" Afribary. Afribary, 19 May. 2021, https://track.afribary.com/works/evaluation-of-conventional-x-ray-machines-in-the-primary-health-care-centers-in-khartoum-state. Accessed 23 Nov. 2024.
Rahman, Amal . "Evaluation Of Conventional X-Ray Machines In The Primary Health Care Centers In Khartoum State". Afribary, Afribary, 19 May. 2021. Web. 23 Nov. 2024. < https://track.afribary.com/works/evaluation-of-conventional-x-ray-machines-in-the-primary-health-care-centers-in-khartoum-state >.
Rahman, Amal . "Evaluation Of Conventional X-Ray Machines In The Primary Health Care Centers In Khartoum State" Afribary (2021). Accessed November 23, 2024. https://track.afribary.com/works/evaluation-of-conventional-x-ray-machines-in-the-primary-health-care-centers-in-khartoum-state